Optitrack calibration

From Rsewiki
Revision as of 16:01, 24 August 2022 by Jca (Talk | contribs)

(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

back to ASTA motion capture

Calibration scale results

The test moves a marker a distance of 1.5m and this should be the distance Optitrack should show. The system is just calibrated, there is perfect coverage.

0000mm.jpg The first position of the test marker.

Distance-0mm.png Optitrack reference detection

1500mm.jpg The end position of the test marker after movement of 1500mm.

Distance-1500mm.png Optitrack distance measurement of the 1500mm movement.

The actual movement starts at about 0.3mm and ends at 1500.6mm - a distance of 1500.3mm with an accuracy of less than 0.5mm.

The optitrack say 1503.97mm a scale error of about 0.3%

May be a problem with the calibration wand.

Personal tools
Namespaces

Variants
Actions
Navigation
Toolbox